Thermo Electron Corporation - - Customer Download
Thermo Fisher Scientific Home  >>
   
 
Customer Manuals
 
Advanced Mass Spectrometry
 
The VIP (Virtual Instrument Partner) Program
 
 
 
   
   
 
 
 
Download - Detail

 

Model: ChromQuest 5.0
Title Hot Fix for ChromQuest 5.0 to Allow Prep-Only Methods for the Accela AS
Description: In ChromQuest 5.0 the prep only box would not remain checked if you tried to save a prep only method using an Accela AS. With this fix installed prep only methods will be allowed.



To install the fix, reboot the computer and copy the files with the ocx and mda extensions into the ChromQuest installation directory. After copying the files, please execute "Register_SurveyorAS_OCX.bat" in the command prompt to register SurveyorAS.ocx. This hot fix has been tested to verify that it corrects the problem, but has not been fully regression tested and should be applied only to systems experiencing these specific problems. If you are not severely affected by these specific problems, you may wish to wait for the next ChromQuest service pack that will contain the fix.

Date: 5/21/2009
File Type: Temp Fixes
File Size: 250KB
Download Now: /peg/CQ5_HF01.zip
   
 